Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 141

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

Module
After you have ...
Loading pump
Replaced the entire pump head assembly
Column
Replaced the rotor seal of the left valve
compartment
Column
Replaced the rotor seal of the right valve
compartment
Column
Replaced the left valve
compartment
Column
Replaced the right valve
compartment
Column
Serviced the instrument (for example, annual
compartment
maintenance)
Column
Performed qualification
compartment
These commands reset the counters and update the information when the action was
performed.
For a list of the commands and counters that are supported for the module, see the
Chromeleon Help.
Control Panel
On the control panel for the module, click Wellness, Qualification, and Service to see the
related predictive performance commands and parameters on separate panels. On these
panels, you can enter the limits and reset the counters. In addition, wellness bars provide
visual indicators of qualification and service periods. The color-coding of the wellness bars
provides information about the status:
Color
Description
Green
OK.
Yellow
The value will soon reach the specified limit and/or the related component needs servicing
or should be replaced soon.
Orange
(Only for monitoring Qualification properties.) The value has reached the specified limit.
However, a Grace Period has been specified during which the pump may still be operated.
Red
The value has reached the specified limit or the specified grace period has expired;
replacement of a component, servicing, or qualification of the pump is overdue. The pump
can no longer be operated; besides, it is not possible to start a batch.
In addition, a message appears in the Chromeleon Audit Trail when a limit has been
reached.
Operating Instructions
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Perform the following
command ...
CheckValveServiceDone,
PistonsChanged, SealChanged,
SupportRingChanged
LeftRotorSealChanged
RightRotorSealChanged
LeftValveChanged
RightValveChanged
ServiceDone
QualificationDone