Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 181

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

The Pressure Transducer Test must be performed when the following components have
been replaced:
♦ Pump head
♦ Flow meter
♦ Flow selector
As long as the retention times are stable and in the expected range, you should not
interrupt pump operation by recalibrating the pressure transducers.
You should check also the offset of the pressure transducers:
♦ Before you test the NC pump for leakage by running the diagnostics functions from
Chromeleon. Observe the recommended order of tests (→ page 134).
♦ If you observe a shift in peaks or if the gradients are not correct.
Operating Instructions
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page 173