5.5.10 Adjusting the Screen Brightness or Contrast

You can adjust the screen brightness or screen contrast to your requirements from
Chromeleon or on the pump display. To do so, select one of the following alternatives:
In Chromeleon, open the Commands dialog box for the pump module. Change the
screen brightness under Brightness and/or the screen contrast under Contrast.
On the pump display, select the Configuration menu (→ page 103) and select Display
& Soft Keys. Change the screen brightness under Brightness and/or the screen contrast
under Contrast.

5.5.11 Operating the Degasser of a SRD-3x00 Solvent Rack

If a SRD-3x00 Solvent Rack is connected to the module, you can operate the degasser of
the solvent rack from the pump module. If the SRD is connected to the module and the
module is powered up, the SRD is powered from the module. The degasser is ready for
operation.
To turn the degasser on and off from the pump display
Select the Preferences menu (→ page 102) and set Degasser to On or Off.
To turn the degasser on and off from Chromeleon
1. In the Server Configuration program, open the Properties dialog for the module
(→ page 53).
2. On the Pumps page, verify that Degasser Control is set to External (→ page 47).
3. Open the Commands dialog box for the pump module and set Degasser to On or Off.
—or—
Open the Tabset Panel for the module (→ page 90) and click More Options (the button
on the utmost left under Commands). In the Degasser group, set Mode to On or Off.
When the degasser is operated in Chromeleon, monitoring of the degasser vacuum and leak
detection is performed from the module. The related commands and properties are
available in the Commands dialog box for the pump module.
Chromeleon Property
DegasserVacuum
SolventRackLeak
Operating Instructions
Description
Reports whether the degasser has reached the operating vacuum.
Reports whether the leak sensor in the SRD detected a leak.