5.3.2 Direct Control

With direct control, you select operating parameters and commands in the Commands
(F8) dialog box. Direct commands are executed as soon as they are entered. For routine
operation, most parameters and commands are available also on a control panel.
To open the Commands dialog box for the module
1. Open a control panel (any panel is possible). To open a control panel, open the
Chromeleon Browser and double-click a control panel in the Dionex Templates/Panels
folder.
2. Connect the control panel to the timebase in which the module is installed. On the
Control menu, select Connect to Timebase, and then select the timebase on the
Timebase tab. For more information about the Timebase dialog, click Help.
Tip:
The Control menu is visible only when a control panel is already open.
3. Press the F8 key or select Command on the Control menu.
4. Show the parameters and commands for the components of the module. They appear
under the names specified for the components in the Properties dialog for the module
(→ page 47).
a) Click the plus sign next to PumpModule. Here you see
♦ General properties and commands related to the entire pump module, such as the
commands and properties for the seal wash system, leak detection, or the
degasser in a SRD-3x00 Solvent Rack that is connected to the module.
♦ Specific groups of properties and commands for the pumps of the pump
module.
Operating Instructions
Fig. 44: Commands dialog box (pump module)
NCS-3500RS and NCP-3200RS
The commands and properties for the entire
pump module appear under PumpModule.
The commands and properties for the
pumps of the pump module appear under
the pump names.