4 Preparation for Operation (Startup)

4.1 Overview

Important:
Important:
After you have unpacked, positioned and connected the module (→ sections 3.1 through
3.4, page 37 and following pages), prepare the module for operation:
1. If the RSLCnano system includes a SRD-3x00 Solvent Rack
Connect the 15-pin D-Sub port on the rear of the Solvent Rack with the Solvent Rack
port on the rear of the module (→ page 25).
2. If you want to connect a mass spectrometer to the module
Connect the Digital I/O port on the rear of the module with the appropriate port on the
mass spectrometer (→ page 42).
3. Connect the solvent reservoirs (→ page 62).
4. Connect drain tubing (→ page 65).
Important: If drain tubing is not connected, modules that are located below this
Important: Si le drainage n'est pas attaché, il peut résulter des dommages par
5. If you want to operate the module from Chromeleon
♦ Set up the module in Chromeleon if it is not already set up (→ page 43).
♦ If you want to operate the degasser of a SRD-3x00 Solvent Rack from Chromeleon,
verify on the Pump page of the Properties dialog for the module that Degasser
Control is set to External (→ page 47).
6. Turn on the power to the module (→ page 85).
Operating Instructions
The module is filled with 2-propanol when being shipped from the
factory. During initial operation of the module, make sure that the
solvents used are miscible with 2-propanol. Otherwise, use an
appropriate intermediate solvent.
L'instrument est stocké sous 2-propanol lorsqu'el est expédié depuis
l'usine. Lors du démarrage initial de l'instrument, assurez-vous que
les solvants utilisés sont miscibles. Dans le cas contraire, utilisez un
solvant intermédiaire approprié. Même d'infimes particules peuvent
endommager le système.
module in the UltiMate 3000 system stack, may suffer severe
damage from the liquid leaving the drain port.
des solvants aux modules qui se trouvent au-dessous de
l'instrument dans l'UltiMate 3000 système.
UltiMate 3000 Series: