6. Connect the other end of the elbow tubing to the pump head inlet.
Removing Solvent Shut Off Valves
1. Stop the pump flow and wait until the pressure is down to zero.
2. Close the shut off valve.
3. Disconnect the elbow tubing from the pump head and empty the line into waste.
4. Unscrew the cap of the solvent reservoir and remove the solvent line together with the
cap from the reservoir.
5. Empty the solvent line to the reservoir into waste.
6. Disconnect the shut off valve with the elbow tubing from the solvent line to the
reservoir.
7. Connect the solvent line between pump head inlet and solvent reservoir.
Operating Instructions
No.
1
2
3
Fig. 80: Installing a shut off valve in a solvent line
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Description
Elbow tubing to pump head
Shut off valve
Solvent line to solvent reservoir
Page 207