UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
9.2 Guidelines on Operating the ProFlow Flow Meter
For operation of the ProFlow flow meter, make yourself familiar with the following
guidelines:
•
Solvent viscosity values that are saved for a Classic flow meter do not apply for the
ProFlow flow meter. A list of commonly used solvents is available in the ProFlow flow
meter. Unknown solvents need to be calibrated newly for the ProFlow flow meter. See
section 5.5.3.1, page 114.
•
When purging the flow meter, mind the different purge times for the ProFlow flow meter
(→ section 4.5.3.1.2, page 75).
•
The ProFlow flow meter includes a built-in inline filter, of which you can replace the
filter frit. For replacement instructions, see section 7.7.3, page 205.
•
Different diagnostic and calibration procedures apply for the ProFlow flow meter. See
section 5.7.2, page 134.
For details on the operating principle of the NC pump with a ProFlow flow meter, see
•
section 8.1.3, page 222.
Page 234
Operating Instructions