Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 209

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

Recommended
Run the General Leak Test from Chromeleon (→ page 134).
In rare cases, new seals may still show an increased leakage rate after several hours of
operation.
If leakage is observed with new piston seals, operate the pump for at least 2 hours at
35 MPa to run in the seals. (It does not help if you tighten the pump head screw more.)
Tip: If the piston seals do not yet seal sufficiently tight, try the following: Have
the pump build up high pressure (→ page 211) and hold the pressure for
some minutes.
Operating Instructions
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page 201