Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 89

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

5. If you want to use column identification
Attach the column ID chip card to the column. Two chip cards are provided in the
accessories kit of the module. Wrap the ribbon around the column, pass the shank of the
rivet through a hole, and press down to unite the two parts.
6. Press the column into the bottom part of the clip and reinstall the ring.
7. Connect the capillaries to the column and establish the fluid connections between the
separation column and the column switching valve.
For information about how to connect the capillaries to the valve, see (→ page 80). When
connecting the capillaries, observe the general information about how to connect and
route the capillaries (→ pages 57 and 60).
8. Activate column identification if applicable (→ page 129).
Operating Instructions
Fig. 40: Attaching the column ID to the column
Fig. 41: Installing a column in the clip
UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page 81