Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 78

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Setting the Solvent Type (ProFlow Flow Meter)
To set the solvent type for the solvent that you are using, follow these steps:
1. In Chromeleon, open the Commands dialog box for the NC pump (→ page 89).
2. Select %A_Solvent (or %B_Solvent) and then select the preferred solvent from the
list.
Tip: If you are using a solvent with up to 2 % additive, it is sufficient to select
the main solvent.
Solvent mixtures with higher percentages of a component require a solvent
calibration.
If the solvent type that you are using is not yet available in the list, perform a solvent
calibration for the solvent to be defined in the flow meter and added to the list in
Chromeleon. See section 5.5.3.1, page 114.
Page 70
Fig. 31: Setting the solvent type (ProFlow flow meter only)
Operating Instructions