Place the capillaries in such a manner that they do not open a small path for ambient air
into the column chamber (that is, place the capillaries preferably in a 90-degree angle
related to the door seal). An improper seal may reduce the heating performance of the
column compartment.
Clips to secure any capillaries routed from the top to the bottom of the column chamber
are provided in the accessories kit of the NCS-3500RS. Insert the capillary into the left or
right side of the clip and attach the clip to the enclosure bottom (→ Fig. 23).
Fig. 23: Clip with capillary attached in the column compartment
Operating Instructions