Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 240

Nano/cap system. nano/cap pump
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
4. Assemble and install the solvent lines with shut off valves. For instructions, see
section 7.8, page 206.
Tip: With the ProFlow flow meter, shut off valves are required during solvent
calibration.
The shut off valves allow you to shut off the solvent flow through the
system at the solvent reservoir, for example, to prevent solvent from
flowing through the system when the pump flow is zero for a longer
period.
5. On the NC pump, replace the capillaries with the capillaries from the Upgrade kit.
This includes:
♦ Capillaries from working cylinder to equilibration cylinder (U-tube) on both pump
heads
♦ Capillary from right pump head to flow meter inlet
♦ Capillary from left pump head to flow meter inlet
For an overview of the connections on the ProFlow flow meter, see section 2.8.2 Flow
Meter, page 30.
Tip: The capillaries can also be used with a Classic flow meter.
6. Connect the module in Chromeleon.
7. Make sure that the firmware version that is installed on the module is suitable for the
ProFlow flow meter. If required, update the firmware to the firmware shipped in the
new Chromeleon version. See section 7.14, page 217.
8. Open the Chromeleon Server Configuration program and adapt the configuration for
your module with the ProFlow flow meter.
Note the following:
♦ If you set up a new system configuration, set up your configuration as required.
♦ If you import an existing system configuration being set up with an older
Chromeleon version, remove the device from the timebase and add it newly to have
the new configuration features available.
a) On the Flowmeter page of the module, verfiy that the Flowmeter Type is set to
ProFlow. Solvent lists show the available pre-defined solvents in the flow meter.
b) On the Signals page of the module, verify that the signal channels for the ProFlow
flow meter are selected. See section 5.5.6, page 121.
Page 232
Operating Instructions