4
Repairing the Detector

Removing the Optical Unit (G1315A/G1365A)

Use this procedure for the the G1315A/G1365A only. For G1315B/G1365B detectors refer
N O T E
to
When required
Tools required
Parts required
Use this procedure for the G1315A/G1365A only.
N O T E
1 Turn off the lamp.
2 Switch off detector, and disconnect power cable.
3 Remove detector from stack and place it on working bench.
4 Remove front cover, top cover and top foam section, see
5 Remove both lamps from the optical unit, see
6 Remove flow cell, see
7 Remove flow cell cover, see
154
"Removing the Optical Unit (G1315B/G1365B)"
For all repairs inside the optical unit
Screwdriver POZI 1 PT3
Optical unit G1315-69002 (exchange assembly) or
Individual parts depending on the following procedures
Cover and Foam"
on page 140.
page 102 (G1315A/G1365A only!).
on page 127 (G1315A/G1365A only!).
Filter"
on page 156.
"Exchanging a Flow Cell"
"Cleaning or Exchanging the Holmium Oxide
1100 Series DAD and MWD Reference Manual
"Removing the Top
"Exchanging a Lamp"
on page 105.
on