Manually Setting The Gnss Position - Spectracom SecureSync 1200 User Reference Manual

Time & frequency reference system
Table of Contents

Advertisement

Sanitization
The Delete Position command must also be used when sanitizing a SecureSync unit (ensuring
that no trace of position data remains on the unit). See "Sanitizing the Unit" on page 312.
Deleting the GNSS position
To delete the GNSS position:
1.
2.
3.
4.
3.3.2.9

Manually Setting the GNSS Position

The exact geographic position (location and elevation) of the antenna your SecureSync unit—
and thus its onboard GNSS receiver—is a major factor for SecureSync to calculate an accur
ate System Time from the GNSS reference.
CHAPTER
3
SecureSync User Reference Guide Rev. 23
Disconnect the GNSS antenna from the SecureSync unit (required only when sanitizing
the unit).
Navigate to Interfaces > References: GNSS Reference . 
Click on the GEAR button next to the GNSS Reference (typically, there is only one ref
erence, numbered "0"). The GNSS 0 window will open:
Locate the Delete Position box, check it, and click Submit.
SecureSync will initiate a GNSS self survey.
Note:
In
Mobile Receiver Mode
the GNSS survey. This feature is only available in
Satellite Mode
(see
"Selecting a GNSS Receiver Mode" on
Satellite Mode a GNSS survey may take up to 24 hours.
Note:
The elevation (altitude) should be set in accordance with the World Geo
detic System 1984 (
WGS 84
it is NOT possible to delete the position and start
Standard Mode
), not Mean Sea Level (MSL).
3.3  Managing References
and in
Single
page 172). In Single