3Z4L Laser Micrometer Protocol
Statistic Processing Non-execution (3000-series)
(Sequence No. 417)
This sequence turns the statistic processing LED off and does not carry out the statistic processing.
Send Data Word Allocation (2nd Operand of PMCR)
None.
Receive Data Word Allocation (3rd Operand of PMCR)
None.
All Statistic Memory Clear (3000-series) (Sequence No. 418)
This sequence clears statistic processing memories of all programs.
Send Data Word Allocation (2nd Operand of PMCR)
None.
Receive Data Word Allocation (3rd Operand of PMCR)
None.
Statistic Processing Memory Clear (3000-series)
(Sequence No. 419)
This sequence clears statistic processing memories of the program under display.
Send Data Word Allocation (2nd Operand of PMCR)
None.
Receive Data Word Allocation (3rd Operand of PMCR)
None.
Statistic Result Request (3000-series) (Sequence No. 420)
This sequence requests the statistic processing results.
Send Data Word Allocation (2nd Operand of PMCR)
None.
Receive Data Word Allocation (3rd Operand of PMCR)
Receive data
storage words
+0
Number of receive data words
(Undefined)
+1
+2
+3
(Undefined)
+4
Decimal portion
+5
Integer portion
(Undefined)
+6
Sign
Decimal portion
+7
Integer portion
+8
(Undefined)
Sign
+9
Decimal portion
+10
+11
(Undefined)
+12
Sign
+13
Decimal portion
+14
(Undefined)
+15
Sign
+16
Decimal portion
Integer portion
+17
(Undefined)
+18
Sign
Program number (P)
Number of statistic data (N)
Average value (AVG)
(Undefined)
Maximum value (MAX)
(Undefined)
Minimum value (MIN)
Integer portion
(Undefined)
Range (R)
Integer portion
(Undefined)
Standard deviation (SD)
(Undefined)
Appendix K
335