Resolution Commands - Agilent Technologies 86038A User Manual

Optical dispersion analyzer
Table of Contents

Advertisement

Base Commands
RFModulationFrequency Property
ValidRFModulationFrequency Method
Agilent 86038A Optical Dispersion Analyzer, Third Edition

Resolution Commands

The Resolution property of the remote client object returns an object,
which handles the resolution of the measurement.
Specifies the RF modulation frequency used to modulate the optical signal
in the ODA test set.
In general, a high modulation frequency minimizes instrumentation noise
and a low modulation frequency maximizes wavelength resolution.
Property Value
Double - RF Modulation Frequency in MHz
Related Command
Resolution.ValidRFModulationFrequency method
VB.NET Syntax
odaClient.Resolution.RFModulationFrequency = 2000
VB 6.0 Syntax
odaClient.Resolution.RFModulationFrequency = 2000
Returns valid minimum, maximum, and default RF modulation frequency
values for the measurement.
Parameters
ByRef Minimum As Double - Minimum RF Modulation Frequency allowed,
in MHz
ByRef Maximum As Double - Maximum RF Modulation Frequency allowed,
in MHz
ByRef Default Value As Double - Default RF Modulation Frequency used, in
MHz.
Related Command
Resolution.RFModulationFrequency property