Repairing the Detector

Exchanging the Power Supply

When required
Tools required
Parts required
The repair level of the power supply assembly is exchanging the complete assembly. No
N O T E
serviceable parts are inside.
1 Turn off the lamp.
2 Switch off the detector, and disconnect the cables.
3 Remove the detector from the stack and place it on the working bench.
4 Remove the front cover, top cover and top foam section, see
5 Remove the detector main board, see
6 Remove the fan assembly, see
7 Remove the optical unit, see
164
If defective
Screwdriver POZI 1 PT3
Wrench 1/4 inch
Wrench 5 mm
Wrench 7 mm
Power supply 0950-2528
Top Cover and Foam"
Board"
on page 143.
page 148 or
"Replacing Fan Parts (G1315B/G1365B)"
(G1315A/G1365A)"
(G1315B/G1365B)"
on page 156.
on page 140.
"Exchanging the Detector Main
"Exchanging the Fan (G1315A/G1365A)"
"Removing the Optical Unit
on page 154 or
"Removing the Optical Unit
1100 Series DAD and MWD Reference Manual
"Removing the
on
on page 150.