DD+DIS033.07E
3.3

Test images

Purpose
The test images are used to check the complete printer function as well as checking
the image quality and certain adjustments
The list enclosed gives an overview of the different test images. This list is based on
software release 1.8.
Name / Purpose
'STI1'
Composed of many different test
images to check all aspects of
image quality.
'STI2'
Composed of many different test
images to check all aspects of
image quality.
'SMPTE'
Standard test image for evaluation
of image quality.
'Medical_image'
Test image to check, whether
stripes or dots, which are slightly
visible in one of the technical
images, are also visible in a
medical test image.
'LinCheck'
Test image with 11 increasing grey
values. Used for density
reproduction check.
'medicalImage' (Mammo)
Test image to check the effect of
possible stripes or dots in mammo
images.