DD+DIS033.07E
HC303002 : upperpickupmodule.film.homeposition
Type
Context
Action
Reason
Cure
HC303003 : upperpickupmodule.film.vacuum
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Part
HC303004 : upperpickupmodule.filmpickup
Type
Context
Action
Reason
Cure
HC303005 : upperpickupmodule.s18.low
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Accidental
None
Upper pickup homing failed accidently
Self resolving accident
Accidental
None
Upper input jam
Remove upper input jam
vacuum system failing
clean vacuum tubes , filter , vacuum cups ......
Lower pickup module defect
Replace
vacuum pump
Accidental
None
Upper input jam
Remove upper input jam
Accidental
None
Upper input jam
Remove upper input jam
DOCUMENT CONTROL NOTE: