Advertisement

DD+DIS032.07E
5.2
Reset of Maintenance Action counters and relative Counters in the
Infocounter
(1)
In the Browser menu select "Service Engineer Tools - Enter Service Command
Session - 7 Service Actions - 4 Maintenance - 4 Clear relative counters".
(2)
Select ' 7 Service Actions - 4 Maintenance - 1 Maintenance Settings'
(3)
Select 'y' for yes at each desired 'Reset Item'. See
(4)
Select "Accept Changes". See
(5)
When leaving the Service menu select "Preventive maintenance" as service
action.
Maintenance Action
Thermal head cleaning required
Replace thermal head anti dust filter
Replace thermal head (end of life)
Film calibration upper tray required
Film calibration lower tray required
Service maintenance required
536709fk.cdr
5.3
Customer Conversation
(1)
Fill out the Maintenance checklist and hand it out to the customer.
(2)
Explain the results of the maintenance to the customer.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
09-2007
Service - Service actions
2
Accept Changes
Discard Changes
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
2
in Figure 23.
Welcome to this
DryStar Axys printer
Factory
Cycles
to go
settings
0
5000
19030
25000
52520
62500
1898
5000
2505
5000
0
5000
Reset Values
Figure 23
Type 5367
Maintenance
1
in Figure 23.
Active
Reset Item
Current
settings
Y
Y
5000
N
N
Y
Y
25000
N
N
Y
Y
62500
N
N
Y
Y
5000
N
N
Y
Y
5000
N
N
Y
5000
Y
N
N
Chapter 9 / 26