DD+DIS033.07E
HC308019 : devilmodule.sn.mismatch
Type
Context
Action
Reason
Cure
HC308020 : devilmodule.power.sensor
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
HC308021 : devilmodule.p42.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E308019
SeeManual
SN mismatch
Call service
Defect
E308020
SeeManual
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
PSU module defect
Replace
MOD09
Defect
E308021
SeeManual
Connector P42 not connected
Connect P42
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08