AGFA DRYSTAR AXYS Service Manual page 220

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC318020 : devilmodule.tyndall.ram
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
HC312025 : devilmodule.tyndall.clock
Type
Context
Action
Reason
Cure
Part
MODULE : powersupplymodule
Elabel : Yes
HC309000 : powersupplymodule.temperature
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
18
Replace
RAM defect
Replace
MOD18
MOD12 cpu board (Thanos or Tyndall) defect
Replace
MOD12
Defect
12
Replace
MOD12 cpu board (Thanos or Tyndall) defect
Replace
MOD12
Defect
E309000
SeeManual
PSU too hot
Switch off printer and let it cool down
PSU module defect
Replace
MOD09
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 54 of 72