AGFA DRYSTAR AXYS Service Manual page 43

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
NOTE:
It is possible to enter the site data "offline" via the site data tool and to enter the site
data via CF card during execution of the printer installation wizard.
Proceeding for preparation of CF card with site data:
(1)
Download the Site Data Tool from
MEDNET, GSO Library:
Hardcopy => Dry Imagers =>
DRYSTAR AXYS => Freeware.
(2)
Install it on the service PC.
(3)
Start the Site Data Tool.
(4)
Enter the Site Data.
(5)
Select "Save":
A zip file is created with the site
data.
(6)
Copy the ZIP file to CF card.
(7)
Enter the CF card when prompted
during execution of the printer
installation wizard. See Figure 1.
NOTE:
As an alternative the site data can also be entered via the printer browser pages in
the key-operator or service menu. Refer to section 3.7.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Figure 2
Chapter 1 / 7
Agfa Company Confidential