DD+DIS033.07E
3.3

Power Supply

LED
Overtemp
+ 3.3 V
+ 5 V
+12 V
-12 V
+24 V
13..27 V
*When the temperature inside the power supply exceeds the permissible limit, printing
can continue for up to 120 seconds. After this delay the power supply switches off the
power and checks the temperature every 20 seconds. As soon as the temperature is
below the permissible limit, it switches on again.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
Electrical and Mechanical Codes, Fuses, LEDs
536634ab.cdr
Status
Color
On
Red
Off
On
Green
Off
On
Green
Off
On
Green
Off
On
Green
Off
On
Green
Off
On
Green
Off
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Figure 3
Function
Temperature inside power supply
exceeded permissible limits. Operation not
possible*.
Temperature inside power supply
within permissible limit.
+3.3 V present
+3.3 V not present
+5 V present
+5 V not present
+12 V present
+12 V not present
-12 V present
-12 V not present
+24 V present
+24 V not present
voltage between 13 V and 27 V
voltage < 13 V
Chapter 3.4 / 9