AGFA DRYSTAR AXYS Service Manual page 212

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC308016 : devilmodule.headvoltage
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Reason4
Cure4
Part
Reason5
Cure5
Part
Reason6
Cure6
Part
HC308017 : duplicate.ipaddress
Type
Context
Action
Reason
Cure
HC308018 : disk.space
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E308016
SeeManual
Connector P1 not connected on TH
Connect P1 on TH
Connector P43 not connected on MOD08 baseboard (Devil or Devius)
Connect P43 on MOD08 baseboard (Devil or Devius)
Connector P3 not connected on MOD08 baseboard (Devil or Devius)
Connect P3 on PSU
PSU module defect
Replace
MOD09
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
TH module defect
Replace
MOD05
Failure
W308017
None
IP adress already in use
Change IP adress
Failure
W308018
None
Harddisk full
Call service