Testimage Lincheck - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.3.5

Testimage LinCheck

Test image to check density reproduction.
Purpose
(2)
In the browser pages select menu
Usage
'Service Engineer Tools - Enter
Service Command Session - 3
Print Testimage – 1 Test Image -
LinCheck.ZIP'.
(3)
The test image as displayed in
Figure 14 will be printed.
(4)
The test image contains a grey
Evaluation
wedge. This grey wedge is
measured during the density
reproduction check.
Refer to chapter 3.6, Adjustments.
NOTE:
For check of density reproduction only measure in the density measurement area, as
there also the CDM is measuring.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Figure 14
Density measurement area
Figure 15
Agfa Company Confidential
536632ga.cdr
536632gb.cdr
Chapter 3.2 / 22