Film Calibration
WHEN TO BE
The film calibration has to be performed in following cases:
PERFORMED
•
Before Smart TH Profile Calibration. See section 2.4.
•
During the compliance film calibration. See section 2.5.
•
After TH replacement.
•
At least each 5000 films (adjustable maintenance action. Refer to chapter 9
Maintenance).
•
In case of a drift of the maximum density.
•
During the troubleshooting procedure.
PURPOSE
Determination of the correlation between control of the thermal head and
corresponding densities on film.
REQUIRED TOOLS:
None
REQUIRED TIME:
10 min.
NOTE:
The film calibration is by default executed automatically at every 5th new film package:
This automatic film calibration can be activated/deactivated in the Key-operator menu
"1. Change settings - 5. Image quality - 1. General - select default for film calibration
(default = ON)". Also the calibration frequency (Default = every 5 film packs) can be
entered there.