AGFA DRYSTAR AXYS Service Manual page 53

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
IMPORTANT:
After a film calibration check the CDM measurement area (see
lightbox: It may not show any dust stripes.
In case dust stripes are visible repeat the film calibration.
NOTE:
The film calibration is successful, if the measured Dmax is < 2% different from the
target Dmax.
The target Dmax depends on the default densitometer which is configured in Key-
operator menu "2 Change Settings - 4 Image Quality - 1 General".
Refer to table below:
Default
Densitometer
X-Rite 310
X-Rite 331
X-Rite 341
X-Rite 301
MacBeth
Gretag D200
When the Dmax isn't reached, the user is informed via local display to retry the
film calibration. The TH power will be increased so that the Dmax will be
reached.
If Dmax is still more than ± 2% different from the target Dmax the display will
show "calibration failed". In this case check the CDM measurement area at the
lightbox for dust stripes. Refer to
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections, and Setup Procedures
Target Dmax DT2B
3.20 O.D.
3.09 O.D.
3.05 O.D.
3.10 O.D.
3.10 O.D.
3.05 O.D.
Figure 11
X
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Figure 11
X
Target Dmax DT2C
3.05 O.D.
2.93 O.D.
2.89 O.D.
2.88 O.D.
2.91 O.D.
2.88 O.D.
.
X
Agfa Company Confidential
) at the
X
Target Dmax DT2
Mammo
3.80 O.D.
3.68 O.D.
3.65 O.D.
3.63 O.D.
3.68 O.D.
3.63 O.D.
Chapter 1 / 17