AGFA DRYSTAR AXYS Service Manual page 96

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
Errors are internally divided into:
• Events
• Warnings
• Defects
In general events are user clearable, warnings are "key-operator"
(= experienced user or hospital technician) clearable, whereas defects have to be
repaired by service personnel.
Warnings and error messages are written into the infocounter.
Diagnosis
everything
yes
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2009
Errors: Classification and treatment
Filmrun
Step 1
Online
event
no
warning
okay?
defect
Filmrun
Step n
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
clearable
Message in plain text.
E.g. 'close upper cover'
key-operator
Message in plain text:
E.g. CDM not calibrated
clearable
Error message:
HC10907:
Suspension motor M3 not
connected or malfunctioning
Figure 24
Type 5367/100
Functional Description
user
clearable
Infocounter
service
536602af.cdr
Agfa Company Confidentia
Chapter 2 / 26

Advertisement

Table of Contents
loading