Mammography Specific Image Adjustments - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
5.4.1

Mammography specific Image Adjustments

NOTE:
By default the Mammography printing functionality of the printer is switched off.
Two conditions have to be fulfilled to allow printing of Mammography images:
Mammo license is activated. See section 3.6.
A DICOM host profile for the Mammo modality with setting "Mammo modality
= yes" has to be created. This 2nd step is described here.
In case the printer has no Mammo modality as input, skip this step.
(1) In the browser pages select "Service Engineer Tools - DICOM profile editor -
New".
(2) Enter Nickname (A).
(3) Enter modality AE title (B).
(4) Select "Mammo Modality = yes" (D).
(5) If a small clear border is required (See Figure 17 next page)
- adjust "image orientation" to "r 180" (C)
- switch on Mammo CWR Rotation* (E).
(6) In case the modality and printer is part of an Agfa Solution, make adjustments as
defined in the solution requirements.
*
CWR = Chest Wall Rotation
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections and Setup Procedure
HELP
Host Profile: CR QS Mammo
A
Nick Name:
B
Use this profile only for incoming 'Calling AE title':
...........................................................
...........................................................
C
Image orientation:
...........................................................
...........................................................
D
Mammo modality:
E
Mammo CWR Rotation:
OK Cancel Show Defaults
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
CR QS Mammo
r180
Yes
No
Show last Input
Figure 16
QS
536701gk.cdr
Chapter 1 / 24