DD+DIS033.07E
HC301020 : uppertraymodule.s11.flagproblem
Type
Context
Action
Reason
Cure
Part
HC301021 : uppertraymodule.s24.flagproblem
Type
Context
Action
Reason
Cure
Part
HC301022 : uppertraymodule.sensors
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC301023 : backside.jam.upper
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W301020
None
S11 defect
Replace
MOD01
Failure
W301021
None
S24 defect
Replace
MOD01
Failure
W301022
None
one of the takeoverdetection sensors is not functioning
Remove jam
Upper tray module defect
Replace
MOD01
Accidental
None
backside jam while printing from upper tray
Insert filmpack correctly
DOCUMENT CONTROL NOTE: