NOTE:
To have minimal clear borders at other mammography printers (DRYSTAR 550X;
DRYSTAR 4500 M), it is not required to set host profile parameter "Rotation" to "r180".
There it is sufficient to set parameters "Mammo CWR rotation" to "YES".
Symptom 2
After adding some modalities via browser key-operator menu it is not possible to print
DICOM images. Following modalities are affected:
Phillips MXVIEW
Phillips MX8000
Fuji ADR series
Fuji Eyepix
Adding these modalities manually via DICOM host profile does not give any problem.
Software problem in software < 1.6.2_C1
Cause
3.2

New Features

Software patch 1.6.2_C1 contains following new features:
Compliance film calibration. Details see 3.2.1.
Activation of a new 10x12" layout for GE Modalities via Browser Service Menu.
Details see 3.2.2.
Possibility to read out future ISO-Format RF-Tags. Details see 3.2.3.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
03-2011
Software Releases and Patches
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Chapter 3.8 / 9
Agfa Company Confidential