Setting Behavior For Unreadable Rf Tags - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.5

Setting Behavior for unreadable RF Tags

Accessible via ...
Accessible via menu:
Purpose of the setting:
Default value:
Example browser
screen:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections, and Setup Procedures
Browser only
Service menu - 2 Change settings - 2 Input Tray
Define the behavior of the printer in case the RF tag is not
readable or available.
Printing not possible
Service - Change input tray settings
If RF-tag is missing or malfunctioning
printing with Dmax = 2.00
printing not possible
Upper input tray
Film view
Tag film format
Tag film type
Lower input tray
Film view
Tag film format
Tag film type
Accept changes Discard changes Reset Values
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Welcome to this
DryStar Axys printer
Normal
Back
14x17"
DT2B
Normal
Back
14x17"
DT2B