AGFA DRYSTAR AXYS Service Manual page 120

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.2.4
How to evaluate the Infocounter
What to check in the
infocounter
1.1 Serial number and
Installation date
1.4 Software Info
What to check in the
infocounter
2.3 Reliability
3.3 Hardware Modification
History
3.4 Software modification
history
5.3 Retries
5.7 Error History
5.8 Error list (sorted by
relative) and
5.9 Error list (sorted by
total):
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
Infocounters.txt evaluation
Comment
Compare device serial number with chapter 8
'Manufacturing Standard Modification' to determine
whether the device possibly requires a modification.
Refer to 3.3 (infocounter.txt), 'HW Modification History', to
check already made modifications.
Compare the currently installed SW version with the latest
SW version described in chapter 3.8, SW releases,
patches. In case a recommended SW patch is available,
install this SW patch.
Refer to section 4.1
Infocounters.txt evaluation
Comment
Should normally be not higher than 0.05. Higher fault
rates need a thorough investigation.
By comparing the status of the device with the available
modifications (refer to Technical Documentation chapter
8) the exact hardware status can be determined.
By checking the software modification history it can be
determined, whether a recent software upgrade solved a
problem, that occurred quite often in the error list in the
past
Many retries (> 1%, based on relative prints) have to be
investigated: They may reduce the throughput of the
device.
Check which errors occurred in the last days, how often
they appeared as well as the CBF (cycles between
failures) of these errors. This gives an overview of the
current status of the machine.
Troubleshoot these errors with the help of the technical
documentation, chapter 3.3, 'Troubleshooting'.
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 14

Advertisement

Table of Contents
loading

Table of Contents