Cleaning of Power Supply Air Inlet

(1)
Clean the power supply air inlet with a
vacuum cleaner.
3.7

Check and Cleaning of Drum

(1)
Open the top cover
(2)
Open the "Hold down bracket"
(3)
Open the TH Module
(4)
Remove the flat belt
(5)
Check the drum for dirt or damage by
turning at the drum drive pulley.
(6)
In case signs of damage are visible,
also check the STI2 test image printed
under step "2.3 Image Quality Check":
A damage will be visible in the flatfield
area of the STI2 test image in
distance of 111 mm.
(7)
Clean the rubber of the drum with a
lint free cloth lightly moistened with
Isopropyl alcohol or Ethanol
(8)
Remount the flat belt