DD+DIS032.07E
Maintenance Checklist
Work Instruction for order no.
The maintenance has to be carried out:
Every 12 months or 25 000 printed films
NOTE:
Maintenance must be carried out according to the maintenance instructions
DD+DIS032.07E
Diagnostics
Questioning of the
customer
Infocounter
analysis
Image quality
check
Maintenance points - Power off
Check of general
condition
Check of condition
of belts and gears
Visual check of
the cables
Cleaning of the
suction cups
Cleaning of
Electronics rack
Cleaning of Power
supply air inlet
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
09-2007
• Questioning of the customer about problems with
the printer.
• Read-out and analyzation of the infocounter
• Printing of STI1, STI2 and medical test image for
comparison with test images at final system check
• Visual check of power and network cable.
• Visual check of covers and film magazines
• Visual check of all belts and gears in the printer.
• Check of cables inside the printer for damage
• Cleaning of the suction cups of both film pickup
assemblies with a lint free cloth lightly moistened
with Isopropyl alcohol
• Cleaning of the Electronic rack with a vacuum
cleaner.
• Cleaning of the Power supply air inlet with a
vacuum cleaner