DD+DIS033.07E
(2)
Field 2 consists of 256 small squares, with grey values from 0 to 255.
Each square has inside another smaller square, with a grey level which is 2
grey steps darker than the surrounding square.
Check: all 256 squares must be visible. In addition, a contrast difference must
be visible between the individual square and the small square inside.
(3)
Field 3 contains a flat field with density value 2.0.
Check: no stripes in any direction may be visible.
NOTE:
In case stripes, dots or insufficient density reproduction in the image can be observed,
refer to the troubleshooting guide (see MEDNET: GSO Library > Hardcopy > Dry
Printers > Drystar Axys > Troubleshooting Guide) for symptom, reason and cure.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007