Defect
05
Replace
TH data is not loaded properly
Replace
MOD05
Defect
E305012
SeeManual
TH has the wrong resolution for this printer OR this software is installed on
the wrong printer
Replace
MOD05
Defect
05
Replace
Connector P3 on PSU not connected
Connect P3 on PSU
Connectors P2 on TH not connected
Connect connector P2
Connector P1 on TH not connected
Connect connector P1 on TH
Connector P43 on MOD08 baseboard (Devil or Devius) not connected
Connect connector P43 on MOD08 baseboard (Devil or Devius)
TH defect
Replace
MOD05