AGFA DRYSTAR AXYS Service Manual page 190

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC304023 : lowerpickupmodule.s16
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC304024 : lowerpickupmodule.s17
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC304025 : lowerpickupmodule.s19
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W304023
None
Lower pickup home sensor S16 malfunctioning
check lower pickup home sensor S16
Lower pickup module defect
Replace
MOD04
Failure
W304024
None
Lower pickup pickup sensor S17 malfunctioning
check lower pickup pickup sensor S17
Lower pickup module defect
Replace
MOD04
Failure
W304025
None
Lower pickup passed sensor S19 malfunctioning
check lower pickup passed sensor S19
Lower pickup module defect
Replace
MOD04