Setting Film Calibration Parameters - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.1

Setting Film Calibration Parameters

Accessible via ...
Accessible via menu:
Purpose of the setting:
Default value:
Example browser
screen:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections, and Setup Procedures
Local keypad and browser
Key-operator menu - 2 Change settings - 4 Image Quality -
1 General
A) Define whether automatic calibration shall be executed
at insertion of a new film pack.
B) If set on ON, define the number of filmpacks, after
which the film calibration shall be executed.
ON - every 5 filmpacks
Key-Operator - Change Image quality settings
Default film calibration
Default Densitometer
Accept changes Discard changes Reset Values
Key-Operator - Change Image quality settings
Select frequency
Accept changes Discard changes Reset Values
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Welcome to this
DryStar Axys printer
off
on
Macbeth TR924
Gretag
X-Rite 301
X-Rite 310
Welcome to this
DryStar Axys printer
Every
packs
5
Figure 3
Agfa Company Confidential
536701fa.cdr
Chapter 1 / 9

Advertisement

Table of Contents
loading