NOTE:
A copy of the infocounter data is always stored in the battery RAM of the printer. At
software installation the contents of the battery RAM are written to the hard disk drive.
This guarantees, that the infocounter data are not lost after exchange of the hard disk
drive
Three possibilities exist to get access to the infocounter (see also Figure 6):
#
Acces to the infocoeunter ...
A
Via printed infocounter on film
B
Via Printers Infocounters Report
C
Via Service Command session
NOTE:
We recommend to use for Infocounter Analysis method A, 'Via printed infocounter on
film', as this is the fastest way to access the infocounter.
3.2.1

Infocounter Analysis via printed infocounter on Film

Required Tools
none
On the local keypad select key-operator menu "Service Actions - Print Infocounters": A
film will be printed with the infocounter data
Analyze the infocounter as described in section 3.2.4
NOTE:
The infocounter is best readable if printed on a 14x17" film.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Refer to ...
3.2.1
3.2.2
3.2.3
Agfa Company Confidential
Chapter 3.2 / 11