DD+DIS033.07E
(1)
In the browser screen select menu 'Service Engineer Tools - Enter Service
Verification:
Command Session - 3 Print Image - 1 Testimage – upper or lower tray* -
STI2.ZIP': The STI2 test image will be printed.
(2)
From the same menu print the medical_image.zip*
*IMPORTANT: Select the tray with 14" width films.
(3)
Check both images at the light box for dark stripes in the area of the 10" film
edge:
• On the STI2 test image dark stripes may be visible in the area of the 10"
• On the medical test image no dark stripes may be visible in the area of the
Result:
The smart profile calibration is performed.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
film edge.
10" film edge.