Testimage Medicalimage 'Mammo; Show Error Codes - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.3.6

Testimage medicalimage 'Mammo'

Test image to check, whether print stripes or dots, which are slightly visible in one of
Purpose
the technical images, are also visible in a medical Mammo test image.
(1)
In the browser pages select menu
Usage
'Service Engineer Tools - Enter
Service Command Session - 3
Print Testimage - 4 medicalimage'.
(2)
Select the tray with Mammo DT2
films.
(3)
The test image as displayed in
Figure 16 will be printed.
(1)
Check whether stripes or dots, that are visible in the STI1 or STI2 image, are
Evaluation
also visible in the testimage 'medicalimage'.
IMPORTANT:
To make this test film available at a printer with software 1.6.0_C1 proceed as
follows:
Connect via telnet to the printer
(1)
At the C:\> prompt enter:
copy C:\images\service\*m.zip D:\images\service
(2)
Print the image from Service Command Session menu - "3 Print Testimage -
4 medicalimage.zip"
3.4

Show Error Codes

View 'Reason - Meaning - Cure' at the local display for an error that appeared at
Purpose
the printer.
Possibility 1: View the last error code:
Usage
(1)
In the key-operator menu select '7 Service Actions - 1 Error messages'
Possibility 2: View all error codes:
(1)
In the browser pages select menu 'Service Engineer Tools - List of errors'
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Figure 16
Agfa Company Confidential
Chapter 3.2 / 23