DD+DIS033.07E
IMPORTANT:
If the printer has 14x11" DT2 Mammo films loaded, select the tray with 14" film for
print head calibration.
If only 8x10" or 10x12" DT2 Mammo films are available, use the tray with 10" films
loaded.
If the printer uses 14" DT2B or DT2B films and 10" DT2 Mammo films, perform the
print head calibration first for the upper tray, then for the lower tray.
Select a 14" film for optimum quality
Field 1
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Welcome to this
DryStar Axys printer
Service - Calibration
1.
Film Calibration
2.
Internal Densitometer Calibration
3.
Thermal Print Head Calibration
4.
Horizontal Alignment
5.
Smart profile calibration
Go back
Welcome to this
DryStar Axys printer
Service - Calibration
Printhead Calibration
Upper Tray
Lower Tray
536736ag.cdr
Field 21
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Adjustments and Calibrations
Welcome to this
DryStar Axys printer
Service - Calibration
Thermal Printhead Calibration
Select Densitometer
Macbeth TR924
Gretag
X-Rite 301
X-Rite 310
Start Calibration
Go back
Welcome to this
DryStar Axys printer
Succeeded
Enter below the measurements of the printer calibration sheet
Step
2.0
00
2.0
02
2.1
04
2.1
06
2.2
08
2.1
10
2.1
12
2.1
14
2.1
16
2.1
18
2.1
20
Accept changes
Discard changes
Figure 15
Reset Values
Step
2.1
01
2.1
03
2.1
05
2.1
07
2.1
09
2.2
11
2.1
13
2.0
15
2.1
17
2.1
19
2.2
21
Reset values