AGFA DRYSTAR AXYS Service Manual page 257

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
INSTRUCTIONS
Via Printer Browser User Interface
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Via Local Keypad
(1)
(2)
(3)
(4)
(5)
(6)
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
In the printer browser user interface select 'Service Engineer Tools'.
Select 'Enter Service Command Session'.
Select '6 Calibration - 4 Horizontal Alignment'.
Select 'upper tray' or 'lower tray': A test film as shown in Figure 1 will be
printed.
Measure the left and right blank border of the test film.
Enter the values in mm (typically approx. 10 mm) in the browser screen.
Select 'Accept Changes'.
In the key-operator menu select menu '7 Service Actions'.
Select '6 Calibration - 4 Horizontal Alignment'.
Select 'upper tray' or 'lower tray': A test film as shown in Figure 1 will be
printed.
Measure the left and right blank border of the test film.
Enter the values in mm (typically approx. 10 mm) in the browser screen.
Select 'confirm'.
DOCUMENT CONTROL NOTE:
Adjustments and Calibrations
DRYSTAR AXYS
Type 5367/100
Repair and Service
Chapter 3.6 / 5