DD+DIS033.07E
HC303010 : upperpickupmodule.homeposition
Type
Context
Action
Reason
Cure
HC303011 : upperpickupmodule.thpositioning
Type
Context
Action
Reason
Cure
HC303012 : upperpickupmodule.elabel.present
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC303013 : upperpickupmodule.elabel.config
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W303010
None
Upper pickup homing failed accidently
Self resolving accident
Accidental
None
Upper input jam
Remove upper input jam
Failure
W303012
None
Upper pickup elabel not connected
Connect upper pickup elabel
Upper pickup Elabel defect
Replace
MOD03
Failure
W303013
None
Upper pickup elabel defect
Replace
MOD03