Error Catalogue
The error catalogue is part of the device software (file "error.xml").
The print-out supplied here is based on software release 2.1.0_C2
It is also available via browser pages, menu 'Service Engineer tools - List of Errors'
IMPORTANT:
Error codes EXXXXX and WXXXXX (displayed on the local display), are registered in the
infocounter as error codes HCXXXXX. The next pages list the HC codes totether with reason
and cure.
Error codes "MOD XX - Please replace module" are also mapped to an HC code. This HC code
can be looked up in key-operator menu "7 Service Actions - Error Codes".
Depending on attributes for "Type" and "Action" (see error descriptions below) error messages
will be issued as error code or plain text:
Type
Defect
Defect
Failure
Accidental
List of Modules:
MOD Nr..
MOD01
MOD02
MOD03 / MOD04
MOD05
MOD06
MOD07
MOD08
MOD09
MOD10
MOD11
MOD12
MOD13
MOD14
MOD15
MOD16
MOD17
MOD18
MOD20
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Action
Error message will be issued as ......
Replace
Message "MODXX Please replace module"
See manual
Message "EXXXXXX – texttext"
None
Message "WXXXXXX - texttext"
None
Message "texttext" (e.g. remove film jam)
Spare Part
UPPER TRAY W/O TAG READER
LOWER TRAY W/O TAG READER
UPPER / LOWER PICK UP MODULE
THERMAL HEAD
OUTPUT MODULE
DRUM DRIVE MODULE
DEVIL MODULE or DEVIUS MODULE
POWER SUPPLY
KEYPAD/DISPLAY MODULE
INPUT SENSORS MODULE
THANOS W/O SDRAM or TYNDALL W/O SDRAM
CAM DRIVE MODULE
GEAR BOX & BELT
SUSPENSION DRIVE MODULE
TPH COOLING FAN
PUMP & VALVES ASSY
RAM MODULE
VILIS COOLING FAN
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting
Chapter 3.3 / 5 of 72