Electrical Check after Power Supply Replacements

In case of power supply replacements or maintenance check the protective earth
resistance according to the national regulations.
The electrical check is split up in two steps:
Step
1
2
1.3.2.1
Basic Protective Earth Test
(1)
Use a multimeter that is capable to
measure a resistance < 1 Ohm
(2)
Put one probe at the ground
connector of the mains input
(3)
Put the other probe at a few metallic
parts, the customer could touch.
The electrical resistance has to be ≤
0.2 Ohm.
Refer to Figure 2.