Printer Bootup-Message - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.6.2

Printer Bootup-Message

The printer terminal level provides following service action:
Purpose
An online logging, which displays all actions of the printer software during
bootup.
The printer terminal level is not intended to be used as official service tool.
It is only used for troubleshooting, e.g. to view the boot-up messages.
(1)
Connect the Service PC to the
Usage
printer as displayed in Figure
41.
(2)
Start the HyperTerminal as
displayed in section 3.6.2.1.
(3)
Switch the printer on:
Now the bootup message
appears
(4)
When the Printer login appears
the bootup message is finished.
3.6.2.1
Setup of the HyperTerminal
NOTE:
The hyper terminal setup as displayed here can be used for all AGFA devices.
Instead of the HyperTerminal following other programs can be recommended to
connect to terminal level:
xt.
xt.exe is an AGFA program, which is part of the IMOS package. It can be started
in a DOS box by typing 'xt'
IMOS Start Menu.
The IMOS Start Menu software has a terminal built in. Select 'tools - terminal'.
Depending on the configuration of IMOS Start, either a DOS box with xt will be
opened or a Hyper terminal window will pop-up.
(1)
To start the setup of the HyperTerminal, go to 'Programs - Accessories -
HyperTerminal - HyperTerminal'
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Figure 41
Agfa Company Confidential
536632ep.cdr
Chapter 3.2 / 45

Advertisement

Table of Contents
loading

Table of Contents