AGFA DRYSTAR AXYS Service Manual page 166

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033
.07E
NOTE:
The printer "Emergency_Install" software can be obtained via MEDNET, GSO library:
http://docs.agfanet/mednetcso-library. Follow path "Hardcopy – Dry Imagers – Drystar
AXYS – Freeware"
CAUTION:
Depending on the software patch version it might be possible, that one of the
boards gets re-programmed during the software upgrade.
In case this re-programming is interrupted by power loss, this specific board
may become unusable.
Do not reboot the printer before requested by a local keypad message
NOTE:
When you choose during Emergency SW Installation to preserve the configuration
following steps are executed internally:
1. A backup of the printer configuration data is created and copied to RAM
2. The software is copied from CF card to hard disk and inflated
3. The boards are programmed
4. The backup is restored from RAM
When you choose during Emergency SW Installation not to preserve the configuration
the configuration data are erased. The printer in this case is in the same state like it
has been shipped from factory
(default IP address 169.254.10.10)
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 60